IVDR TDF QC Check: Information Supplied with the Device
Review the information supplied with the device (label, packaging and instructions for use) in the technical documentation against Regulation (EU) 2017/746 (IVDR), Annex I, Chapter III, Section 20 and the applicable references. Mark each check OK, Issue or N/A. Issues are collected into a report that is emailed to the QC reviewer, consultant and team lead.
1. Project
2. Checklist
0/0 reviewed · 0 issuesSet sterile supply, device type, IFU format and intended user above to show only the relevant sections.
Common NB findings: check these first
0/7- N1
IFU intended purpose covers every IVDR element: what is detected or measured (analyte / marker), function (screening, diagnosis, monitoring, prognosis, prediction, companion diagnostic), condition, specimen type, testing population, intended user, automated or not, qualitative / semi-quantitative / quantitative
- N2
Analytical and clinical performance claims in the IFU match the performance evaluation report exactly; no claim without supporting data
- N3
Electronic IFU used only for devices intended for professional use, never for self-testing or near-patient testing (Reg. (EU) 2021/2226 is an MDR regulation and does not apply to IVDs)
- N4
Self-testing / near-patient testing: label and IFU wording validated with representative intended users; result reading and follow-up actions are understood
- N5
Reagents containing dangerous substances or mixtures carry CLP pictograms, signal words and hazard / precautionary statements; SDS available unless all information is in the IFU
- N6
Metrological traceability of calibrator and trueness-control values stated, with reference materials / reference measurement procedures of higher order
- N7
IFU carries the serious-incident reporting notice (to the manufacturer and the Competent Authority of the Member State where the user and/or patient is established)
A. Regulatory basis and applicable references
0/6- A1
Device qualifies as an IVD under Art. 2(2); risk class (A, B, C, D) and classification rule recorded with rationale
- A2
Applicability of Annex I §20.1, 20.2, 20.3, 20.4.1 and 20.4.2 mapped clause by clause, with a rationale for every N/A clause
- A3
TD contains the complete set of labels and IFUs: outer, inner and primary container labels, all kit components and all languages for the intended markets
- A4
Basic UDI-DI and UDI-DI assigned; UDI carrier applied on the label and on all higher packaging levels
- A5
Manufacturer and device data on the label (name, SRN, address, Basic UDI-DI) match the EUDAMED registration
- A6
CE marking applied per Art. 18 and Annex V, with NB number for Class B, C, D and Class A sterile; consistent with the EU DoC
B. Harmonised standards, common specifications and guidance
0/8- B1
Standards list identifies each labelling standard with edition and IVDR harmonisation status; approach to standards follows MDCG 2021-5
- B2
EN ISO 18113-1 (terms, definitions and general requirements) applied; current edition and harmonisation status checked
- B3
Reagents: EN ISO 18113-2 (professional use) or EN ISO 18113-4 (self-testing) applied as relevant
- B4
Instruments: EN ISO 18113-3 (professional use) or EN ISO 18113-5 (self-testing) applied as relevant
- B5
EN ISO 15223-1 applied for symbols, including the IVD symbol; current edition, amendments and harmonised listing checked
- B6
EN ISO 14971 used to cross-check warnings, precautions, limitations and residual risks
- B7
Instrument / software standards applied for markings and IFU content (e.g. EN IEC 61010-1 and -2-101, EN IEC 61326-2-6, EN 62304, EN IEC 82304-1)
- B8
Class D: applicable common specifications (Reg. (EU) 2022/1107) checked for performance and labelling-related requirements
C. General requirements (Annex I §20.1)
0/10- C1
Medium, format, content, legibility and location suit the device, its intended purpose and the intended user's knowledge; drawings / diagrams used where helpful
- C2
Required particulars appear on the device itself or, where not practicable (e.g. small vials), on the packaging for each unit (primary / inner) and on the outer packaging
- C3
Label is human-readable; machine-readable forms (barcode, RFID) only supplement it
- C4
No IFU / abbreviated IFU only in duly justified, exceptional cases where safe and intended use without an IFU is demonstrated
- C5
IFU supplied in paper by default; electronic IFU only for professional-use devices (not near-patient testing), with label stating how to access it
- C6
Self-testing and near-patient testing devices: IFU supplied in paper (hard copy) with each device
- C7
Single IFU copy for multiple devices to one user / location only where agreed with the purchaser; further copies free on request
- C8
Residual risks to be communicated are included as limitations, contra-indications, precautions or warnings
- C9
Symbols from EN ISO 15223-1 used; any symbol or identification colour not covered by a harmonised standard or CS is described in the documentation supplied
- C10
Label and IFU in the official Union language(s) required by each target Member State, under a controlled translation procedure
D. Outer packaging and kit label (Annex I §20.2)
0/16- D1
Device name / trade name and catalogue, model or reference number; intended purpose where not obvious
- D2
Legal manufacturer name / registered trade name or trade mark and registered place of business
- D3
EC REP name and address shown for manufacturers based outside the EU
- D4
Clear IVD designation: statement "In vitro diagnostic medical device" or the IVD symbol (EN ISO 15223-1); or, where applicable, device for performance study
- D5
Total kit contents indicated: volume, weight, number of tests or items included
- D6
Lot number (LOT) or serial number (SN) shown with the ISO 15223-1 symbol
- D7
UDI carrier present in both AIDC (linear / 2D barcode) and HRI form with verified print quality
- D8
Use-by date unambiguous (YYYY-MM-DD or YYYY-MM) and supported by real-time stability data; date of manufacture where no use-by applies
- D9
Storage and handling conditions (temperature limits, humidity, light, other) shown and aligned with stability and transport studies
- D10
Sterile status and sterilisation method declared if supplied sterile, or any special microbial state / cleanliness
- D11
In-use / open-vial stability (after opening, reconstitution, thawing or preparation) shown where different from shelf life, or referenced to the IFU
- D12
Full CLP alignment for hazardous chemicals, preservatives (e.g. sodium azide) or biological hazards: GHS pictograms, signal word, hazard (H) and precautionary (P) statements (detail in IFU where space is insufficient)
- D13
Disclosure where reagents contain material of human or animal origin, including infectious agents or biological control sera (biological risk warning)
- D14
Other warnings / precautions needing the user's immediate attention shown, kept to a minimum with detail in the IFU
- D15
Self-testing or near-patient testing use stated explicitly on the label
- D16
Single-use indication where applicable; kit components also sold separately are labelled as devices in their own right
E. Primary / small component labels (vials, ampoules, cartridges)
0/5- E1
Primary container shows component name and volume / quantity
- E2
Primary container shows lot (LOT) or serial (SN) number
- E3
Primary container shows an unambiguous use-by date
- E4
Primary container shows storage conditions (or symbol)
- E5
Component identity clear: reagent / active ingredient, buffer, calibrator or control material (with level where relevant); hazard information where required
F. Sterile packaging (Annex I §20.3)
0/4- F1
Sterile packaging identified as such, with a declaration of sterile condition and the sterilisation method
- F2
Sterile packaging shows manufacturer name and address, device description, month and year of manufacture and use-by date
- F3
Instruction to check the IFU if the sterile packaging is damaged or unintentionally opened; IFU gives the action to take
- F4
Sterile barrier system and sterilisation method consistent with validation (EN ISO 11607-1/-2, EN ISO 11135 / 11137)
G. IFU: identification, intended purpose and test principle
0/4- G1
Intended purpose stated in full: analyte / marker / parameter, qualitative, semi-quantitative or quantitative, clinical indication and function (screening, confirmatory, monitoring, diagnosis, CDx), patient population and intended user (laboratory professional, trained healthcare worker or lay person)
- G2
Companion diagnostics: INN of the associated medicinal product stated
- G3
Test principle described: scientific principle, reaction or detection method (e.g. PCR, ELISA, lateral flow)
- G4
Metrological traceability of calibrator and control values to reference materials / higher-order reference measurement procedures
H. IFU: reagents, materials and specimens
0/5- H1
Reagent composition: nature and quantity, proportion or concentration of active ingredients, with buffers and other components
- H2
Materials provided vs materials required but not provided (ancillary equipment, reagents, consumables) listed completely
- H3
Preparation and storage: reconstitution, thawing, mixing, dilution; validated in-use and working-solution stability stated (e.g. stable 30 days at 2–8 °C after opening)
- H4
Specimen type(s) stated (e.g. serum, EDTA plasma, nasal swab, capillary whole blood)
- H5
Pre-analytical requirements: patient preparation, collection tube types, anticoagulant compatibility, centrifugation speed / time, transport, storage temperature, specimen stability and allowable freeze–thaw cycles
I. IFU: test procedure, calibration and quality control
0/4- I1
Step-by-step procedure: sample dilution, incubation times / temperatures, washing, read-out window
- I2
Calibration procedure, calibration frequency and zero / blank setting described
- I3
Internal and external QC procedures, acceptance criteria and actions when a control fails stated
- I4
Compatible instruments, devices and general-purpose equipment identified, with any known restrictions on combinations
J. IFU: result interpretation and performance claims
0/8- J1
Criteria for reading results: cut-off values, thresholds, positive / negative / indeterminate (invalid) decision points; whether confirmatory testing is needed
- J2
Mathematical approach / calculation for quantitative results described
- J3
Biological reference intervals for defined populations stated where relevant
- J4
Limit of blank, limit of detection and limit of quantitation stated
- J5
Analytical specificity, cross-reactivity and endogenous / exogenous interferents (haemolysis, icterus, lipaemia, common drugs) stated
- J6
Measuring interval and linearity stated
- J7
Precision stated: repeatability (within-run) and reproducibility (between-day / lot / site / operator); trueness and accuracy where relevant
- J8
Clinical performance: diagnostic sensitivity and specificity, PPV and NPV where relevant, consistent with the CPR
K. IFU: limitations, warnings and regulatory notices
0/6- K1
Procedural limitations stated: interfering substances (HIL), hook / prozone effect limits, false-positive and false-negative scenarios, specimen limitations and populations in which the assay is not validated
- K2
Warnings and precautions: signs of malfunction or degradation, external influences, infectious or hazardous materials, sensitising substances; special facilities or training where relevant
- K3
Safe handling, biological decontamination and waste disposal (biohazard waste, sharps, CLP-classified chemicals; infection, environmental and physical hazards)
- K4
Class C and D: IFU references the location of, and link to, the Summary of Safety and Performance on EUDAMED
- K5
Serious-incident reporting notice to the manufacturer and the Competent Authority
- K6
IFU date of issue / revision date and version number, with modifications clearly indicated; manufacturer contact for technical assistance
L. Instruments, analysers and software
0/15- L1
Intended platform role, compatible assays and automated steps performed (e.g. extraction, amplification, detection)
- L2
Measurement principle described: optical detection system, thermal cycling, fluidics path, sensor mechanics
- L3
Instrument performance stated: precision across runs / wells / positions, carry-over limits, photometric / fluorescence linearity
- L4
Clinical performance cross-referenced to the assay-specific IFUs / application sheets
- L5
Sample handling: rack types, dead-volume requirements, accepted tube types and dimension tolerances
- L6
Electrical safety and EMC disclosures: ratings, voltage, frequency, power consumption, fuse ratings; declarations aligned with EN IEC 61010-1, EN IEC 61010-2-101 and EN IEC 61326-2-6
- L7
Ambient operating limits: temperature, humidity, altitude and ambient light
- L8
Installation verification and readiness checks described
- L9
Interoperability: compatible LIS protocols (e.g. HL7, ASTM / CLSI LIS2), middleware specification, minimum hardware, operating system and IT network requirements
- L10
Data security / cybersecurity measures, including protection against unauthorised access
- L11
Consumables and reagent compatibility: qualified reagents, wash solutions, cuvettes and calibration consumables required for valid results clearly specified
- L12
Calibration: protocol and frequency, factory calibration standards, optical / performance verification tools and start-up self-test diagnostics
- L13
Maintenance: daily / periodic intervals, user-serviceable vs manufacturer-serviceable parts, consumable replacement
- L14
Cleaning and decontamination procedures, including decontamination before service, and biohazard spill management
- L15
Error / fault messages and diagnostic codes listed with corrective actions
M. Self-testing and near-patient testing (Annex I §19, §20.4.2)
0/8- M1
Label and IFU legible and understandable for the intended user, confirmed by usability testing
- M2
Illustrated step-by-step instructions in simple, non-clinical language: reagent preparation, specimen collection, running the test, reading the result
- M3
Visual guide (photos / drawings) for positive, negative and invalid results; results presented so they are readily understood
- M4
Advice on action for positive, negative and indeterminate results; test limitations; possibility of false positive / false negative; factors that may affect the result
- M5
Self-testing: prominent statement not to take any medical decision or adjust medication without consulting a healthcare professional; information on the disease, its prevalence and national advice sources where available
- M6
Self-testing for monitoring a diagnosed condition: patient adapts treatment only after appropriate training
- M7
Self-testing: instructions for safe disposal of used and contaminated components at home (lancets, swabs, cassettes, buffer)
- M8
Near-patient testing: prominent, fast-action warnings and internal / external QC routines suitable for non-laboratory clinical staff (nurses, clinic staff)
O. Consistency with TD, risk, performance evaluation and PMS
0/6- O1
Each warning, precaution, limitation and contra-indication traces to a risk control or residual risk in the risk file
- O2
Claims in label / IFU supported by the PER (scientific validity, analytical and clinical performance)
- O3
Class C and D: Summary of Safety and Performance consistent with the IFU and written for the intended user (and patient where relevant)
- O4
Label and IFU consistent with the DoC, EUDAMED data and marketing material; no misleading claims
- O5
PMS, PMPF, vigilance and FSCA outputs reviewed for labelling impact; label / IFU changes under change control
- O6
Class C and D self-testing: SSP includes a dedicated summary written in language a lay person can understand
P. Reference control and QC sign-off
0/6- P1
Legal IVDR requirements, harmonised standards, common specifications, non-binding guidance and internal I3CGLOBAL QC criteria kept separate
- P2
Not every standard or guidance applied to every device: applicability and exclusion rationale recorded
- P3
Controlled reference register shows title, identifier, edition, publication date, applicability, harmonisation status, link and date last checked
- P4
Current IVDR harmonised standards list, common specifications and latest MDCG revisions verified on the Commission website before final approval
- P5
Each deficiency logged with corrective action, owner, target date and close-out sign-off
- P6
Final review signed by QA reviewer, RA specialist, usability / systems engineer and performance evaluation expert