EU REP vs UK Responsible Person vs Swiss CH-REP: Which Representatives Do You Need?
The EU REP vs UK Responsible Person vs CH-REP question has a simple rule: a medical device or IVD manufacturer needs a local representative in every one of these markets where it is not established. An EU Authorized Representative (EU REP) covers the EU, a UK Responsible Person (UKRP) covers Great Britain, and a Swiss authorised representative (CH-REP) covers Switzerland. This guide compares the three roles and shows which ones your company needs.
EU REP vs UK Responsible Person vs CH-REP: What Is the Difference?
All three are local representatives that act for a manufacturer that is not established in the market. They hold or give access to the technical documentation, act as the contact point for the regulator and pass on complaints and incident reports. The differences are in who must appoint one, how registration works, what goes on the label and how far liability goes.
| Topic | EU REP (EU) | UK Responsible Person (Great Britain) | CH-REP (Switzerland) |
|---|---|---|---|
| Legal basis | MDR and IVDR Article 11 (tasks) and Article 12 (change) | UK Medical Devices Regulations 2002 (as amended) | MedDO Art. 51–52; IvDO Art. 44–45 |
| Who must appoint one | Manufacturers not established in the EU | Manufacturers based outside the UK that place devices on the Great Britain market | Manufacturers not established in Switzerland, including EU and EEA manufacturers |
| Regulator | National competent authority of the EU Member State | MHRA | Swissmedic |
| Registration | Actor registration in EUDAMED, which gives the representative an SRN | Registers the manufacturer’s devices with the MHRA before they are placed on the GB market | Registers with Swissmedic and receives a CHRN |
| Mandate scope | At least all devices of the same generic device group | Written evidence of the manufacturer’s authority to act | At least all devices of the same generic device group |
| Regulatory compliance person | Must have a PRRC available (Article 15(6)) | Not covered by the MHRA guidance we checked | Must have a PRRC (Art. 52(1) MedDO / Art. 45(1) IvDO) |
| Liability for defective devices | Jointly and severally liable with the manufacturer in the cases set out in Article 11(5) | Not covered by the MHRA guidance we checked | The Swissmedic information sheet sets no separate liability rule |
EU REP vs UK Responsible Person: Which Does My Company Need?
Use this EU REP vs UK Responsible Person vs CH-REP table to check your obligations. Find the country where your company is established. “Yes” means you must appoint that representative before placing devices on that market.

| Manufacturer established in | EU REP | Northern Ireland | UKRP (Great Britain) | CH-REP |
|---|---|---|---|---|
| USA, Canada, India, China, Japan, Korea and other non-European countries | Yes | YesAn EU or Northern Ireland representative; one EU REP covers both | Yes | Yes |
| EU Member State | No | No | Yes | Yes |
| Iceland or Norway (EEA) | No | No | Yes | Yes |
| Liechtenstein (EEA) | No | No | Yes | NoCustoms treaty with Switzerland |
| Great Britain (England, Scotland, Wales) | Yes | YesAn EU or Northern Ireland representative | NoRegisters directly with the MHRA | Yes |
| Northern Ireland | No | No | NoUK-based, registers directly | Yes |
| Switzerland | Yes | Yes | Yes | No |
| Turkey | NoEU–Turkey Customs Union | No | Yes | Yes |
Turkey option: according to the European Commission’s notice on the EU–Turkey Customs Union, a manufacturer established outside both the EU and Turkey needs only one authorised representative, located either in the EU or in Turkey, to access both markets.
What Does an EU Authorized Representative Do?
Under MDR and IVDR Article 11, a manufacturer not established in the EU can place a device on the EU market only after designating a sole authorised representative by written mandate. The representative’s minimum tasks are:
- Verify that the EU declaration of conformity and technical documentation exist and that the right conformity assessment was done
- Keep copies of the technical documentation, declaration of conformity and certificates for the required period
- Register in EUDAMED and verify the manufacturer’s UDI and device registration
- Give competent authorities the information they request and cooperate on corrective actions
- Forward complaints and suspected incident reports to the manufacturer
- Terminate the mandate if the manufacturer acts against its obligations, and inform the authority
The European Commission explains these duties in guidance MDCG 2022-16. Read more on our European Authorized Representative page.
What Does a UK Responsible Person Do?
Since 1 January 2021, manufacturers based outside the UK must appoint a UK Responsible Person to place devices on the Great Britain market. This includes EU manufacturers. The MHRA only accepts direct device registrations from UK-based manufacturers. According to the MHRA guidance, the UK Responsible Person must:
- Register the manufacturer’s devices with the MHRA before they are placed on the Great Britain market
- Make sure the declaration of conformity and technical documentation are prepared, and keep them available for MHRA inspection
- Provide information and cooperate with the MHRA on corrective actions
- Pass complaints and incident reports to the manufacturer
- End the relationship and inform the MHRA if the manufacturer breaches its obligations
- Hold written evidence of the manufacturer’s authority to act
Importers and distributors do not need to appoint a UK Responsible Person. Read more on our UK Responsible Person page.
What Does a Swiss Authorised Representative (CH-REP) Do?
Since 26 May 2021, the EU–Swiss mutual recognition agreement no longer covers medical devices, so Swiss manufacturers need an EU REP and EU manufacturers need a CH-REP. Manufacturers not established in Switzerland, including EU and EEA manufacturers, may place devices on the Swiss market only after designating a CH-REP. Manufacturers in Liechtenstein are exempt because of its customs treaty with Switzerland. According to the Swissmedic information sheet MU600_00_016, the CH-REP must:
- Register with Swissmedic and obtain a Swiss Single Registration Number (CHRN)
- Keep the technical documentation available, or agree with the manufacturer that it is provided to Swissmedic within 7 days on request
- Verify that the declaration of conformity, technical documentation and certificates exist
- Have a person responsible for regulatory compliance (PRRC)
- Report serious incidents and safety corrective actions in Switzerland to Swissmedic
- Keep a register of complaints, non-conforming devices, recalls and withdrawals
- From 1 January 2027, verify that the devices are registered in swissdamed
- Keep declarations of conformity and certificates for 10 years, or 15 years for implantable devices, after the last device was placed on the market
I3CGLOBAL does not act as Swiss authorised representative. Contact Swissmedic or a Swiss-based provider for this role.
How Do Registration and Labelling Differ?
EU: EUDAMED and SRN
The EU REP registers as an actor in EUDAMED and receives a Single Registration Number (SRN). Its name and address go on the label, usually with the EC REP symbol from EN ISO 15223-1. See EUDAMED registration.
Great Britain: MHRA registration
The UK Responsible Person registers the manufacturer’s devices with the MHRA. All devices, including IVDs, custom-made devices and system or procedure packs, must be registered before they are placed on the Great Britain market.
Switzerland: CHRN
The CH-REP registers with Swissmedic and receives a CHRN. For MDR devices, the CH-REP must be on the label. For IVDR devices, it goes on the label or an accompanying document, but on the label for self-tests.
Are EU, UK and Swiss Representatives Liable for Defective Devices?
In the EU, MDR and IVDR Article 11(5) make the authorised representative legally liable for defective devices on the same basis as the manufacturer, jointly and severally, if the manufacturer is not established in the EU and has not met its Article 10 obligations. From 9 December 2026, the new EU Product Liability Directive (EU) 2024/2853 also allows claims against the authorised representative of a non-EU manufacturer.
The Swissmedic information sheet sets no separate liability rule for the CH-REP; it lists its tasks and refers to the written mandate. The MHRA guidance we checked does not set out a liability rule for the UK Responsible Person. In all three markets, the written mandate should state clearly who is responsible for what.
Do I Need a Representative for Northern Ireland?
The EU MDR has applied in Northern Ireland since 26 May 2021 and the EU IVDR since 26 May 2022. According to the UK government guidance:
- Manufacturers based in Great Britain must designate an authorised representative based in the EU or Northern Ireland to supply Northern Ireland
- Manufacturers based in Northern Ireland do not need an EU authorised representative
- For manufacturers outside the UK and the EU, one EU REP covers both the EU and Northern Ireland
Is the CE Mark Still Accepted in Great Britain?
Yes. According to the MHRA guidance, Great Britain currently accepts CE marked devices as follows:
| Device type | Accepted in Great Britain until |
|---|---|
| General devices under the EU MDR, and IVDs under the EU IVDR | 30 June 2030 |
| IVDs under the EU IVDD | Certificate expiry or 30 June 2030, whichever is sooner |
| General devices under the EU MDD or AIMDD | Certificate expiry or 30 June 2028, whichever is sooner |
On 21 October 2026, the UK government announced that it will recognise MDR and IVDR CE marks indefinitely and extend recognition of MDD-marked devices to 31 December 2028. New legislation is needed before this takes effect. Even with a CE mark, non-UK manufacturers still need a UK Responsible Person and MHRA registration for Great Britain.
What Are the Most Common Mistakes When Appointing Representatives?
Most problems in the EU REP vs UK Responsible Person vs CH-REP decision come from treating the three markets as one. These are the mistakes we see most often:
- Assuming a CE mark is enough for Great Britain. A valid CE mark is accepted in Great Britain, but non-UK manufacturers still need a UK Responsible Person and MHRA device registration.
- Assuming EU manufacturers need nothing for Switzerland. Since 26 May 2021, EU and EEA manufacturers need a CH-REP to place devices on the Swiss market. Only Liechtenstein manufacturers are exempt.
- Forgetting Northern Ireland. Manufacturers based in Great Britain need an authorised representative in the EU or Northern Ireland to supply Northern Ireland, because the EU MDR and IVDR apply there.
- A mandate that does not cover the whole device group. In the EU and Switzerland, the mandate must cover at least all devices of the same generic device group.
- Labels that do not match the appointment. The EU REP must appear on the label, and for MDR devices so must the CH-REP. Plan label changes before you switch representatives.
- Registering in the wrong order. In the EU, the representative’s EUDAMED actor registration and SRN come before device registration. In Great Britain, devices must be registered with the MHRA before they are placed on the market.
Before you start, list every market you sell in, where your company is established and which device groups you sell. That list tells you exactly which representatives you need and what each mandate must cover.
How Can I3CGLOBAL Help?
I3CGLOBAL is a medical device and IVD regulatory consultancy. We act as EU Authorized Representative from our offices in Germany and Portugal, and as UK Responsible Person for Great Britain. We do not act as Swiss authorised representative.
EU Authorized Representative
Article 11 mandate for all device classes, EUDAMED actor registration and SRN, and the PRRC. EU REP services
UK Responsible Person
MHRA device registration and UK Responsible Person services for Great Britain. UKRP services
Registration support
Support for UDI data and free sale certificates for export markets.
Need an EU REP or UK Responsible Person?
Tell us where your company is established, your device classes and your target markets. We will confirm which representatives you need and send a proposal for the EU and Great Britain.
- EU REP from Germany or Portugal
- UK Responsible Person and MHRA registration
- EUDAMED, SRN and UDI support
Get a proposal
Request an EU REP proposal or check our annual fees.
EU REP vs UK Responsible Person vs CH-REP: Frequently Asked Questions
What is the difference between an EU REP and a UK Responsible Person?
An EU Authorized Representative acts for a non-EU manufacturer on the EU market under MDR and IVDR Article 11 and registers in EUDAMED. A UK Responsible Person acts for a non-UK manufacturer on the Great Britain market and registers the manufacturer’s devices with the MHRA. They are separate roles, and one does not replace the other.
Does an EU manufacturer need a UK Responsible Person?
Yes. Manufacturers based outside the UK, including EU manufacturers, must appoint a UK Responsible Person to place devices on the Great Britain market. The MHRA only accepts direct registrations from UK-based manufacturers.
Does an EU manufacturer need a Swiss authorised representative?
Yes. Since 26 May 2021 the EU–Swiss mutual recognition agreement no longer covers medical devices, so manufacturers not established in Switzerland, including EU and EEA manufacturers, must designate a CH-REP. Manufacturers in Liechtenstein are exempt.
Does a UK manufacturer need an EU Authorized Representative?
Yes. Manufacturers based in Great Britain need an EU REP for the EU market, and an authorised representative based in the EU or Northern Ireland to supply Northern Ireland. Manufacturers based in Northern Ireland do not need an EU authorised representative.
Can one company act as EU REP, UK Responsible Person and CH-REP?
Yes, if it is established in each market and meets each set of duties: an EU address for the EU REP, a UK address for the UK Responsible Person and a Swiss address with a CHRN for the CH-REP. I3CGLOBAL acts as EU REP and UK Responsible Person but not as CH-REP.
Do importers or distributors need a UK Responsible Person?
No. According to MHRA guidance, importers and distributors do not need to appoint a UK Responsible Person. The requirement applies to manufacturers based outside the UK.
What is a CHRN?
A CHRN is the Swiss Single Registration Number that Swissmedic issues to Swiss authorised representatives and importers. A company that is both importer and CH-REP receives two CHRNs. No Swissmedic licence is required.
Is a CE mark enough to sell in Great Britain?
A valid CE mark is currently accepted in Great Britain until the dates set by the MHRA, and the UK government announced on 21 October 2026 that MDR and IVDR CE marks will be recognised indefinitely once new legislation is in place. Non-UK manufacturers still need a UK Responsible Person and MHRA registration.
Do Turkish manufacturers need an EU Authorized Representative?
No. Under the EU–Turkey Customs Union, Turkish manufacturers do not need an EU authorised representative. Manufacturers outside both the EU and Turkey need one authorised representative in either the EU or Turkey.