MDR TDF QC Check: Information Supplied with the Device
Internal · QC DepartmentReview the labelling and IFU section of the MDR technical file against Annex I, Chapter III (GSPR 23) and the applicable references. Mark each check OK, Issue or N/A. Issues are collected into a report that is emailed to the QC reviewer, consultant and team lead.
1. Project
2. Checklist
0/0 reviewed · 0 issuesSet sterile supply, implantable, IFU format and target user above to show only the relevant sections.
Common NB findings: check these first
0/7- N1
IFU carries the serious-incident reporting notice (to the manufacturer and the Competent Authority) verbatim, without softened wording
- N2
Intended purpose, indications, target population and contraindications on label / IFU match the CER and risk file word for word
- N3
UDI carrier present in both AIDC (2D DataMatrix / linear barcode) and HRI form, with verified print grade ≥ 1.5 (C)
- N4
Reusable devices: maximum number of reuses, inspection / end-of-life criteria and validated reprocessing parameters stated in the IFU
- N5
eIFU used only with Reg. (EU) 2021/2226 fulfilled: documented risk assessment, paper copy within 7 calendar days, access details on label
- N6
CMR 1A/1B or endocrine-disrupting substances > 0.1% w/w (e.g. DEHP, DBP) labelled on device / packaging and explained in the IFU
- N7
Every symbol not taken from EN ISO 15223-1 is fully explained in the IFU or accompanying documentation
A. Regulatory basis and applicable references
0/6- A1
Regulatory route confirmed: device is under MDR 2017/745 (IVDs need a separate IVDR checklist, Annex I §20); class and rule recorded
- A2
Applicability of Annex I §23.1, 23.2, 23.3 and 23.4 mapped clause by clause, with a rationale for every N/A clause
- A3
TD contains the complete set of labels and IFUs: all packaging levels and all languages for the intended markets
- A4
Basic UDI-DI and UDI-DI assigned and UDI carrier requirements applied to every packaging level
- A5
Device and manufacturer data on the label (name, SRN, address, Basic UDI-DI) match the EUDAMED registration
- A6
CE marking applied per Art. 20 and Annex V, with NB number where required, consistent with the DoC
B. Harmonised standards and guidance
0/7- B1
Standards list identifies each labelling standard with edition and EU harmonisation status; approach to standards follows MDCG 2021-5
- B2
EN ISO 15223-1:2021 + A1:2025 applied for symbols; current harmonised listing and transition arrangements checked
- B3
ISO 20417 edition confirmed (ISO 20417:2026 published March 2026); presumption of conformity claimed only for a harmonised EN edition
- B4
EN ISO 14971 used to cross-check warnings, precautions, residual risks and user information
- B5
EN ISO 11607-1 / -2 applied to sterile barrier system labelling and packaging validation
- B6
Sterilisation standard matches the declared method (EN ISO 11135 EO, EN ISO 11137 radiation, EN ISO 17665 moist heat)
- B7
Product- or technology-specific standards applied for additional labelling, warnings and symbol requirements (e.g. IEC 60601-1, ISO 80369)
C. General principles and delivery formats (GSPR 23.1)
0/5- C1
Required particulars appear on the device itself, or a documented rationale shows direct marking is not practicable or appropriate
- C2
Lay users: plain wording, legible type size, consult-a-professional circumstances stated, warnings validated in summative usability testing
- C3
All symbols and identification colours conform to EN ISO 15223-1 (proportions, colours) or are described where no standard exists
- C4
No IFU supplied (Class I / IIa): rationale shows safe use without instructions; 23.4(p) and (v) information available on request
- C5
Label and IFU translated into the languages required by each target Member State, under a controlled translation procedure
D. Label (GSPR 23.2)
0/8- D1
Label shows trade name, model / catalogue number, package contents, intended purpose where not obvious, and the MD symbol
- D2
Manufacturer legal name and registered address match the QMS / MDR certificate; EC REP name and address shown for non-EU manufacturers
- D3
CE mark in correct proportions with the valid 4-digit NB number (Class Is, Im, Ir, IIa, IIb, III)
- D4
Lot or serial number shown with the ISO 15223-1 LOT / SN symbol; implantables carry serial (active) or serial / lot number
- D5
Use-by date unambiguous (at least year and month, YYYY-MM-DD or YYYY-MM) or date of manufacture where no use-by applies; matches aging data
- D6
Storage and handling conditions (temperature, humidity, light) listed and aligned with stability and transit testing
- D7
Single-use device shows the do-not-reuse symbol and the IFU explains the risks of reuse, based on the risk file
- D8
Label indicates medicinal substances, human or animal tissues / derivatives, and qualitative composition for substance-based devices
E. Sterile packaging and sterile barrier system (GSPR 23.3)
0/4- E1
Sterile packaging recognisable as such; single or double sterile barrier symbol per EN ISO 11607-1
- E2
Sterile state and method declared on the sterile barrier (STERILE EO, STERILE R, STERILE steam) and on the label
- E3
Instruction to check the IFU if the sterile packaging is damaged or unintentionally opened; IFU gives the action to take
- E4
Sterile packaging shows manufacturer name and address, device description, month and year of manufacture and use-by date
F. Instructions for use (GSPR 23.4)
0/9- F1
IFU states intended purpose, indications, contraindications, patient target group, intended users and use environment
- F2
Clinical benefits and performance characteristics match the CER and PMS data; Class III and implantables link to the SSCP in EUDAMED
- F3
Residual risks, undesirable side-effects, warnings and precautions from the ISO 14971 file are fully listed
- F4
Compatible devices, accessories and software / hardware versions, IT network and IT security requirements specified
- F5
Reusable devices: validated cleaning, disinfection and sterilisation parameters and end-of-life criteria stated
- F6
Installation check, preventive maintenance, consumables, calibration intervals and routine safety checks described
- F7
Radiation-emitting devices: nature, type, intensity and distribution of radiation and protection measures stated
- F8
Safe disposal of device, accessories and consumables: infection, sharps, WEEE and hazardous components covered
- F9
IFU date of issue, or date and identifier of the latest revision, clearly shown
G. Electronic IFU (Reg. (EU) 2021/2226)
0/3- G1
Device and intended users qualify for eIFU under Reg. (EU) 2021/2226 (as amended)
- G2
Documented eIFU risk assessment covers user access, failure modes and emergency information
- G3
Paper IFU supplied free of charge within 7 calendar days on request; label shows eIFU symbol, website and how to request paper
H. Implant card and patient information (Art. 18)
0/4- H1
Art. 18 applicability decided against the exempted device list (Art. 18(3)) and implant card device types
- H2
Implant card gives device name, type, serial / lot number, UDI, manufacturer name, address and website, with fields for patient, date and centre
- H3
Patient information covers expected lifetime, follow-up, warnings and precautions, and qualitative / quantitative material composition
- H4
Card and leaflet state MRI status (MR Safe, MR Conditional, MR Unsafe) with verified conditions
I. Consistency with TD, risk, clinical and PMS
0/4- I1
Each warning, precaution and contraindication traces to a risk control or residual risk ID in the risk file
- I2
Claims and performance characteristics in label / IFU are supported by the CER and V&V data
- I3
Label and IFU consistent with the DoC, EUDAMED data, SSCP and marketing material
- I4
PMS, vigilance and FSCA outputs reviewed for labelling impact; label / IFU changes under change control
J. Reference control and QC sign-off
0/6- J1
Legal MDR requirements, harmonised standards, non-binding guidance and internal I3CGLOBAL QC criteria kept separate
- J2
Not every standard or guidance applied to every device: applicability and exclusion rationale recorded
- J3
Controlled reference register shows title, identifier, edition, publication date, applicability, harmonisation status, link and date last checked
- J4
Current harmonised standards list and latest MDCG revisions verified on the Commission website before final approval
- J5
Each deficiency logged with corrective action, owner, target date and close-out sign-off
- J6
Final review signed by QA reviewer, RA specialist, usability / systems engineer and medical safety / CER author