MDR / IVDR TDF QC Check: Risk Management
Review the risk management file in the technical documentation against EN ISO 14971:2019 + A11:2021, ISO/TR 24971:2020 and Annex I, Chapter I of Regulation (EU) 2017/745 (MDR) or Regulation (EU) 2017/746 (IVDR). Select the regulation in section 1 to show the relevant checks. Mark each check OK, Issue or N/A. Issues are collected into a report that is emailed to the QC reviewer, consultant and team lead.
1. Project
2. Checklist
0/0 reviewed · 0 issuesSet regulation, software and intended user above to show only the relevant checks.
Common NB findings: check these first
0/8- N1
Criteria for risk acceptability defined in the policy and RM plan, consistent with the state of the art and the MDR / IVDR requirement to reduce risks as far as possible (no economic or ALARP-only arguments)
- N2
Hazard identification is complete and systematic (all hazard categories, intended use and reasonably foreseeable misuse), not limited to an FMEA
- N3
Probability and severity estimates justified with data sources (complaints, literature, test data, similar devices), not assumed; uncertainty and limitations of the estimates considered
- N4
Implementation AND effectiveness of every risk control measure verified, with reference to the V&V report
- N5
Information for safety is used only after design and protective measures; residual risks disclosed in IFU / label are traceable to the RMF
- N6
Overall residual risk and benefit-risk conclusion supported by the clinical evaluation (MDR) or performance evaluation (IVDR)
- N7
PMS, complaints, vigilance, PMCF / PMPF and literature data fed back into the RMF, with RMF updated and dated accordingly
- N8
RMF is specific to this device, its intended purpose and its actual design and variants - not a generic or template file; hazards, estimates and controls reference this device's features, materials and data
A. Regulatory basis and risk management system
0/9- A1
Risk management system established, implemented, documented and maintained as a continuous iterative process throughout the entire lifecycle
- A2
RM procedure integrated in the QMS and applied to product realisation
- A3
Top management policy for establishing criteria for risk acceptability documented and approved
- A4
Competence of persons performing RM tasks (incl. clinical / application expertise) documented
- A5
RM file structured and traceable for every identified hazard: analysis, evaluation, control, verification and residual risk
- A6
EN ISO 14971:2019 + A11:2021 applied, including Annex ZA (MDR) and Annex ZB (IVDR) relationship to the GSPRs; ISO/TR 24971:2020 used as guidance
- A7
TD Annex II §5 contains the benefit-risk analysis (Annex I §1 and §8) and the solutions adopted and results of risk management (Annex I §3)
- A8
Risk-related Chapter I GSPRs mapped to the RMF: MDR Annex I §1–§5, §8 (and §9 for Annex XVI products); IVDR Annex I §1–§5, §8
- A9
Supporting standards and guidance listed with editions: EN 62366-1:2015 + A1:2020, EN 62304:2006 + A1:2015, IEC 82304-1, MDCG 2019-16 Rev. 1, MDCG 2020-6 (MDR legacy) / MDCG 2022-2 (IVDR)
B. Risk management plan (EN ISO 14971 §4.4)
0/7- B1
Scope: device / device family, accessories, variants and configurations, intended purpose, intended users and use environment, and the lifecycle phases covered by the plan clearly defined; the plan is specific to this device, not a generic ISO 14971 template
- B2
Responsibilities and authorities assigned
- B3
Requirements for review of RM activities defined
- B4
Criteria for risk acceptability and the methodology for estimating probability and severity defined and justified for this device, before individual risks are evaluated, based on the policy and including criteria where probability cannot be estimated; the risk matrix or other evaluation method is described with meaningful severity categories and consistently defined probability categories (a 5 × 5 matrix is not mandatory - the chosen method and its limitations are justified)
- B5
Method to evaluate overall residual risk and benefit-risk acceptability criteria defined in the plan before the clinical / performance evaluation
- B6
Activities for verification of implementation and effectiveness of risk control measures defined
- B7
Activities for collection and review of production and post-production information defined and linked to the PMS plan, with criteria for reviewing and updating the RMF
C. Risk analysis (EN ISO 14971 §5)
0/11Hazard category reference: categories and examples to consider (C3)
| Hazard category | Examples to consider |
|---|---|
| Biological | Toxicity, sensitisation, irritation, infection and contamination |
| Mechanical | Breakage, fracture, leakage, sharp edges, entrapment and loss of structural integrity |
| Electrical | Electric shock, leakage current, overheating and electrical failure |
| Thermal | Excessive temperature, burns and thermal injury |
| Chemical | Toxic substances, degradation products, leachables and chemical incompatibility |
| Radiation | Unintended radiation exposure, where applicable |
| Usability | Incorrect operation, foreseeable use error and misunderstanding of instructions |
| Software | Incorrect output, algorithm errors, data corruption, loss of functionality and inappropriate clinical decisions |
| Cybersecurity | Unauthorised access, compromised integrity, loss of availability and unsafe software behaviour |
| Infection control | Contamination, inadequate sterilisation, reprocessing failure and packaging breach |
| Performance | Inadequate performance, incorrect results or failure to achieve the intended purpose |
| Environmental | Temperature, humidity, electromagnetic interference, transport and storage conditions |
| Manufacturing | Process deviations, material variability, assembly errors and supplier-related failures |
| Post-market | Previously unidentified hazards, emerging risks, complaints, adverse events and trend information |
- C1
Intended use and reasonably foreseeable misuse documented: intended purpose, indications and contraindications, patient population, intended users (incl. training / competence), use environment, operating principle and essential functions, accessories / compatible devices and interfaces, device lifetime and frequency of use, variants and configurations, and MDR / IVDR class - all consistent with the device description in the TD
- C2
Characteristics related to safety identified (e.g. ISO/TR 24971 Annex A questions answered)
- C3
Known and foreseeable hazards identified across all applicable categories (see the hazard category reference above) and all lifecycle phases, including manufacturing, transport, storage, use and disposal, with sequences of events leading to hazardous situations and harms
- C4
Analysis method(s) defined in the plan, appropriate to the device and its complexity, and consistently applied (PHA, FMEA, FTA, HAZOP, use-related risk analysis); hazards, hazardous situations and harms clearly distinguished; ISO 14971 does not prescribe one universal method or matrix
- C5
MDR: hazards from materials, biocompatibility, sterility, infection, energy, electrical / EMC, mechanical and combination with other devices addressed; implant-related risks for implantable devices and reprocessing risks for reusable devices included where relevant
- C6
IVDR: hazards from incorrect, invalid or delayed results analysed: false positive, false negative, invalid / indeterminate results and failure to produce a result, and their impact on clinical decisions, patients and public health; misleading or misinterpreted results considered even where the result is analytically valid
- C7
Risk estimation: severity ratings justified; probability estimated and justified where feasible (criteria for non-estimable probability applied otherwise); data sources, assumptions, historical data, literature, testing and expert assessment documented
- C8
IVDR: pre-analytical, analytical and post-analytical hazards (sample stability, matrix interference, cross-reactivity, hook effect, carry-over, calibration, reagent stability, result reporting) and user infection / chemical exposure analysed
- C9
Use-related hazards from the usability engineering process included: primary operating functions, use errors, perception / cognition / action hazards, user-interface mitigations
- C10
Software hazards and software item contribution to hazardous situations analysed; IT security risks addressed
- C11
Risks from intended users and training assessed; lay users: incorrect use, misinterpretation of results and delayed medical action analysed
D. Risk evaluation and risk control (EN ISO 14971 §6, §7)
0/8- D1
Each estimated risk evaluated consistently against the acceptability criteria in the RM plan; unacceptable risks identified for further control; no blanket 'acceptable' ratings without criteria, rationale and evidence
- D2
Risk control options applied in the mandatory priority order: inherently safe design and manufacture, protective measures (incl. alarms), information for safety and, where appropriate, training; warnings / IFU never used to avoid feasible design or protective measures, and any reliance on information for safety is justified
- D3
Risks reduced as far as possible without adversely affecting the benefit-risk ratio; no economic argument used to stop risk reduction
- D4
Each risk control measure clearly identified; implementation and effectiveness verified, with evidence traceable to the relevant test, analysis or inspection
- D5
Residual risk evaluated after controls for each hazardous situation
- D6
Where residual risk is not acceptable, further controls applied or a benefit-risk analysis performed and documented; risks that remain unacceptable are addressed, not waived
- D7
New hazards or increased risks introduced by risk control measures assessed and shown not to make risks unacceptable; residual risk reassessed after implementation
- D8
Completeness of risk control confirmed for all identified hazardous situations
E. Overall residual risk and benefit-risk (EN ISO 14971 §8)
0/8- E1
Overall residual risk evaluated using the method and criteria in the RM plan, considering the contribution of all residual risks
- E2
Benefit-risk determination documented in the TD, with benefits linked to the intended purpose, severity and probability of harms considered, and uncertainties and limitations acknowledged; not based solely on an assertion of benefit
- E3
MDR: benefits and benefit-risk conclusion supported by clinical data in the CER; side-effects acceptable against benefits
- E4
IVDR: benefits (clinical / analytical) and benefit-risk conclusion supported by the performance evaluation report
- E5
Significant residual risks disclosed to users, with the information needed to do so included in the IFU / label
- E6
MDR: clinical benefits defined and quantified as patient-relevant outcomes (e.g. survival, symptom reduction, function) with measurable parameters
- E7
IVDR: benefits defined as diagnostic accuracy (sensitivity, specificity, PPV, NPV) and clinical utility / public health impact
- E8
Benefit-risk benchmarked against the state of the art and alternative treatments / diagnostic methods; alternative risk controls considered where appropriate; conclusion consistent with the intended purpose, claims and IFU
F. Risk management review and report (EN ISO 14971 §9)
0/4- F1
RM review performed before commercial release: plan implemented, overall residual risk acceptable, production / post-production methods in place
- F2
RM report confirms completion of all planned activities, documents the required conclusions, addresses outstanding issues and limitations, and is reviewed and approved by the persons with assigned responsibility
- F3
RMF revision, date and device version match the current TD revision and device configuration
- F4
RM report and RMF updated and re-approved when significant new information (PMS, vigilance, design or process changes) requires revision
G. Production and post-production information (EN ISO 14971 §10)
0/5- G1
System for collecting and reviewing production and post-production information (complaints, NC, CAPA, vigilance, literature, similar devices, state of the art) in place
- G2
MDR: PMS / PSUR (Art. 83–86) and PMCF outputs evaluated for impact on hazards, risk estimates and acceptability
- G3
IVDR: PMS / PSUR (Art. 78–81) and PMPF outputs evaluated for impact on hazards, risk estimates and acceptability
- G4
Trend reporting, serious-incident data and relevant field safety corrective actions (own and similar devices) reviewed for new or increased risks
- G5
Design, manufacturing, supplier and labelling changes assessed for risk impact under change control; resulting RMF changes documented and approved
H. Traceability and consistency with the TD
0/6- H1
Each hazard traces to risk controls, verification evidence and residual risk acceptance
- H2
Warnings, precautions, contra-indications and limitations in IFU / label trace to residual risks
- H3
GSPR checklist references the RMF for all applicable risk-related GSPRs
- H4
RMF consistent with design inputs / outputs, design V&V reports, biological evaluation, sterilisation and packaging validation, and electrical safety / EMC testing, where applicable
- H5
RMF consistent with usability file, software V&V and cybersecurity documentation, PMS plan and PMS report / PSUR, and CAPA / complaint-handling records
- H6
Harms, side-effects and their frequencies consistent between RMF and CER / PER and SSCP / SSP
I. Reference control and QC sign-off
0/4- I1
Legal MDR / IVDR requirements, harmonised standards, non-binding guidance and internal I3CGLOBAL QC criteria kept separate
- I2
Current harmonised standards list and latest MDCG revisions verified on the Commission website before final approval
- I3
Each deficiency logged with corrective action, owner, target date and close-out sign-off
- I4
Final review signed by QA reviewer, RA specialist and clinical / performance evaluation expert