Review the risk management file in the technical documentation against EN ISO 14971:2019 + A11:2021, ISO/TR 24971:2020 and Annex I, Chapter I of Regulation (EU) 2017/745 (MDR) or Regulation (EU) 2017/746 (IVDR). Select the regulation in section 1 to show the relevant checks. Mark each check OK, Issue or N/A. Issues are collected into a report that is emailed to the QC reviewer, consultant and team lead.

1. Project

2. Checklist

0/0 reviewed · 0 issues

Set regulation, software and intended user above to show only the relevant checks.

Common NB findings: check these first

0/8
  1. N1

    Criteria for risk acceptability defined in the policy and RM plan, consistent with the state of the art and the MDR / IVDR requirement to reduce risks as far as possible (no economic or ALARP-only arguments)

    Common NB finding · Annex I §2, §3 · EN ISO 14971 §4.2, §4.4 · Annex ZA / ZB · see B4 · Evidence: RM policy + RM plan acceptability matrix

  2. N2

    Hazard identification is complete and systematic (all hazard categories, intended use and reasonably foreseeable misuse), not limited to an FMEA

    Common NB finding · EN ISO 14971 §5.2–5.4 · ISO/TR 24971 Annex A · see C1–C4 · Evidence: Hazard list + PHA / FMEA / FTA

  3. N3

    Probability and severity estimates justified with data sources (complaints, literature, test data, similar devices), not assumed; uncertainty and limitations of the estimates considered

    Common NB finding · EN ISO 14971 §5.5 · see C7 · Evidence: Risk estimation rationale + data sources

  4. N4

    Implementation AND effectiveness of every risk control measure verified, with reference to the V&V report

    Common NB finding · EN ISO 14971 §7.2 · see D4 · Evidence: Risk control traceability to V&V reports

  5. N5

    Information for safety is used only after design and protective measures; residual risks disclosed in IFU / label are traceable to the RMF

    Common NB finding · Annex I §4 · EN ISO 14971 §7.1, §8 · see D2, H2 · Evidence: Risk control option analysis + IFU traceability

  6. N6

    Overall residual risk and benefit-risk conclusion supported by the clinical evaluation (MDR) or performance evaluation (IVDR)

    Common NB finding · Annex I §1, §8 · Art. 61 (MDR) / Art. 56 (IVDR) · see E3, E4 · Evidence: RM report vs CER / PER

  7. N7

    PMS, complaints, vigilance, PMCF / PMPF and literature data fed back into the RMF, with RMF updated and dated accordingly

    Common NB finding · Annex I §3(e), (f) · EN ISO 14971 §10 · see G1–G5 · Evidence: RMF revision history vs PMS / PSUR outputs

  8. N8

    RMF is specific to this device, its intended purpose and its actual design and variants - not a generic or template file; hazards, estimates and controls reference this device's features, materials and data

    Common NB finding · EN ISO 14971 §4.5, §5 · Annex II §1.1 · see C1, C3 · Evidence: RMF content vs device description and design documents

A. Regulatory basis and risk management system

0/9
  1. A1

    Risk management system established, implemented, documented and maintained as a continuous iterative process throughout the entire lifecycle

    MDR / IVDR Annex I §3 · Art. 10(2) · EN ISO 14971 §4.1 · Evidence: RM procedure + RMF

  2. A2

    RM procedure integrated in the QMS and applied to product realisation

    EN ISO 13485 §7.1 · Art. 10 · Annex IX · Evidence: QMS procedure reference

  3. A3

    Top management policy for establishing criteria for risk acceptability documented and approved

    EN ISO 14971 §4.2 · Evidence: RM policy

  4. A4

    Competence of persons performing RM tasks (incl. clinical / application expertise) documented

    EN ISO 14971 §4.3 · Evidence: Training / qualification records

  5. A5

    RM file structured and traceable for every identified hazard: analysis, evaluation, control, verification and residual risk

    EN ISO 14971 §4.5 · Evidence: RMF index + traceability matrix

  6. A6

    EN ISO 14971:2019 + A11:2021 applied, including Annex ZA (MDR) and Annex ZB (IVDR) relationship to the GSPRs; ISO/TR 24971:2020 used as guidance

    Art. 8 · EN ISO 14971 Annex ZA / ZB · ISO/TR 24971 · Evidence: Standards list + GSPR checklist

  7. A7

    TD Annex II §5 contains the benefit-risk analysis (Annex I §1 and §8) and the solutions adopted and results of risk management (Annex I §3)

    Annex II §5 · Evidence: TD section 5 vs RMF and benefit-risk analysis

  8. A8

    Risk-related Chapter I GSPRs mapped to the RMF: MDR Annex I §1–§5, §8 (and §9 for Annex XVI products); IVDR Annex I §1–§5, §8

    MDR / IVDR Annex I Chapter I · Evidence: GSPR checklist rows vs RMF references

  9. A9

    Supporting standards and guidance listed with editions: EN 62366-1:2015 + A1:2020, EN 62304:2006 + A1:2015, IEC 82304-1, MDCG 2019-16 Rev. 1, MDCG 2020-6 (MDR legacy) / MDCG 2022-2 (IVDR)

    Art. 8 · MDCG register · Evidence: Standards and guidance list

B. Risk management plan (EN ISO 14971 §4.4)

0/7
  1. B1

    Scope: device / device family, accessories, variants and configurations, intended purpose, intended users and use environment, and the lifecycle phases covered by the plan clearly defined; the plan is specific to this device, not a generic ISO 14971 template

    EN ISO 14971 §4.4(a) · Evidence: RM plan

  2. B2

    Responsibilities and authorities assigned

    EN ISO 14971 §4.4(b) · Evidence: RM plan

  3. B3

    Requirements for review of RM activities defined

    EN ISO 14971 §4.4(c) · Evidence: RM plan

  4. B4

    Criteria for risk acceptability and the methodology for estimating probability and severity defined and justified for this device, before individual risks are evaluated, based on the policy and including criteria where probability cannot be estimated; the risk matrix or other evaluation method is described with meaningful severity categories and consistently defined probability categories (a 5 × 5 matrix is not mandatory - the chosen method and its limitations are justified)

    EN ISO 14971 §4.4(d) · Annex I §2 · Evidence: RM plan acceptability matrix

  5. B5

    Method to evaluate overall residual risk and benefit-risk acceptability criteria defined in the plan before the clinical / performance evaluation

    EN ISO 14971 §4.4(e) · ISO/TR 24971 §8 · Evidence: RM plan

  6. B6

    Activities for verification of implementation and effectiveness of risk control measures defined

    EN ISO 14971 §4.4(f) · Evidence: RM plan

  7. B7

    Activities for collection and review of production and post-production information defined and linked to the PMS plan, with criteria for reviewing and updating the RMF

    EN ISO 14971 §4.4(g) · Annex III · Evidence: RM plan vs PMS plan

C. Risk analysis (EN ISO 14971 §5)

0/11
Hazard category reference: categories and examples to consider (C3)
Hazard categoryExamples to consider
BiologicalToxicity, sensitisation, irritation, infection and contamination
MechanicalBreakage, fracture, leakage, sharp edges, entrapment and loss of structural integrity
ElectricalElectric shock, leakage current, overheating and electrical failure
ThermalExcessive temperature, burns and thermal injury
ChemicalToxic substances, degradation products, leachables and chemical incompatibility
RadiationUnintended radiation exposure, where applicable
UsabilityIncorrect operation, foreseeable use error and misunderstanding of instructions
SoftwareIncorrect output, algorithm errors, data corruption, loss of functionality and inappropriate clinical decisions
CybersecurityUnauthorised access, compromised integrity, loss of availability and unsafe software behaviour
Infection controlContamination, inadequate sterilisation, reprocessing failure and packaging breach
PerformanceInadequate performance, incorrect results or failure to achieve the intended purpose
EnvironmentalTemperature, humidity, electromagnetic interference, transport and storage conditions
ManufacturingProcess deviations, material variability, assembly errors and supplier-related failures
Post-marketPreviously unidentified hazards, emerging risks, complaints, adverse events and trend information
  1. C1

    Intended use and reasonably foreseeable misuse documented: intended purpose, indications and contraindications, patient population, intended users (incl. training / competence), use environment, operating principle and essential functions, accessories / compatible devices and interfaces, device lifetime and frequency of use, variants and configurations, and MDR / IVDR class - all consistent with the device description in the TD

    EN ISO 14971 §5.2, §5.3 · EN 62366-1 · Annex II §1.1 · Evidence: RMF vs device description

  2. C2

    Characteristics related to safety identified (e.g. ISO/TR 24971 Annex A questions answered)

    EN ISO 14971 §5.3 · ISO/TR 24971 Annex A · Evidence: Safety characteristics list

  3. C3

    Known and foreseeable hazards identified across all applicable categories (see the hazard category reference above) and all lifecycle phases, including manufacturing, transport, storage, use and disposal, with sequences of events leading to hazardous situations and harms

    EN ISO 14971 §5.4 · ISO/TR 24971 Annex A · Evidence: Hazard list / PHA

  4. C4

    Analysis method(s) defined in the plan, appropriate to the device and its complexity, and consistently applied (PHA, FMEA, FTA, HAZOP, use-related risk analysis); hazards, hazardous situations and harms clearly distinguished; ISO 14971 does not prescribe one universal method or matrix

    EN ISO 14971 §5 · ISO/TR 24971 Annex B · Evidence: Analysis records

  5. C5

    MDR: hazards from materials, biocompatibility, sterility, infection, energy, electrical / EMC, mechanical and combination with other devices addressed; implant-related risks for implantable devices and reprocessing risks for reusable devices included where relevant

    MDR Annex I §10–§17 · EN ISO 10993-1 · IEC 60601-1 · Evidence: Hazard list vs GSPR checklist

  6. C6

    IVDR: hazards from incorrect, invalid or delayed results analysed: false positive, false negative, invalid / indeterminate results and failure to produce a result, and their impact on clinical decisions, patients and public health; misleading or misinterpreted results considered even where the result is analytically valid

    IVDR Annex I §1–§4 · ISO/TR 24971 Annex H · MDCG 2022-2 · Evidence: IVD result-error hazard analysis

  7. C7

    Risk estimation: severity ratings justified; probability estimated and justified where feasible (criteria for non-estimable probability applied otherwise); data sources, assumptions, historical data, literature, testing and expert assessment documented

    EN ISO 14971 §5.5 · Evidence: Risk estimation rationale

  8. C8

    IVDR: pre-analytical, analytical and post-analytical hazards (sample stability, matrix interference, cross-reactivity, hook effect, carry-over, calibration, reagent stability, result reporting) and user infection / chemical exposure analysed

    IVDR Annex I · ISO/TR 24971 Annex H · Evidence: IVD hazard analysis

  9. C9

    Use-related hazards from the usability engineering process included: primary operating functions, use errors, perception / cognition / action hazards, user-interface mitigations

    MDR / IVDR Annex I §5 · EN 62366-1:2015 + A1:2020 · Evidence: Use-related risk analysis vs usability file

  10. C10

    Software hazards and software item contribution to hazardous situations analysed; IT security risks addressed

    EN 62304 §7 · EN IEC 81001-5-1 · MDCG 2019-16 · Evidence: Software risk analysis + cybersecurity risk file

  11. C11

    Risks from intended users and training assessed; lay users: incorrect use, misinterpretation of results and delayed medical action analysed

    MDR Annex I §22 / IVDR Annex I §19 · Evidence: Lay-user risk analysis

D. Risk evaluation and risk control (EN ISO 14971 §6, §7)

0/8
  1. D1

    Each estimated risk evaluated consistently against the acceptability criteria in the RM plan; unacceptable risks identified for further control; no blanket 'acceptable' ratings without criteria, rationale and evidence

    EN ISO 14971 §6 · Evidence: Risk evaluation table

  2. D2

    Risk control options applied in the mandatory priority order: inherently safe design and manufacture, protective measures (incl. alarms), information for safety and, where appropriate, training; warnings / IFU never used to avoid feasible design or protective measures, and any reliance on information for safety is justified

    MDR / IVDR Annex I §4 · EN ISO 14971 §7.1 · Evidence: Risk control option analysis

  3. D3

    Risks reduced as far as possible without adversely affecting the benefit-risk ratio; no economic argument used to stop risk reduction

    MDR / IVDR Annex I §2 · EN ISO 14971 Annex ZA / ZB · Evidence: Risk control rationale

  4. D4

    Each risk control measure clearly identified; implementation and effectiveness verified, with evidence traceable to the relevant test, analysis or inspection

    EN ISO 14971 §7.2 · Evidence: V&V reports cross-referenced in RMF

  5. D5

    Residual risk evaluated after controls for each hazardous situation

    EN ISO 14971 §7.3 · Evidence: Residual risk table

  6. D6

    Where residual risk is not acceptable, further controls applied or a benefit-risk analysis performed and documented; risks that remain unacceptable are addressed, not waived

    EN ISO 14971 §7.4 · Evidence: Benefit-risk analysis

  7. D7

    New hazards or increased risks introduced by risk control measures assessed and shown not to make risks unacceptable; residual risk reassessed after implementation

    EN ISO 14971 §7.5 · Evidence: RMF review record

  8. D8

    Completeness of risk control confirmed for all identified hazardous situations

    EN ISO 14971 §7.6 · Evidence: Traceability matrix

E. Overall residual risk and benefit-risk (EN ISO 14971 §8)

0/8
  1. E1

    Overall residual risk evaluated using the method and criteria in the RM plan, considering the contribution of all residual risks

    EN ISO 14971 §8 · ISO/TR 24971 §8 · Evidence: RM report

  2. E2

    Benefit-risk determination documented in the TD, with benefits linked to the intended purpose, severity and probability of harms considered, and uncertainties and limitations acknowledged; not based solely on an assertion of benefit

    Annex II §5 · Annex I §1 · Evidence: Benefit-risk analysis

  3. E3

    MDR: benefits and benefit-risk conclusion supported by clinical data in the CER; side-effects acceptable against benefits

    MDR Art. 61 · Annex XIV · Annex I §8 · MEDDEV 2.7/1 Rev. 4 §9.3 / A7 · Evidence: CER benefit-risk section vs RMF

  4. E4

    IVDR: benefits (clinical / analytical) and benefit-risk conclusion supported by the performance evaluation report

    IVDR Art. 56 · Annex XIII · Annex I §8 · MDCG 2022-2 · Evidence: PER benefit-risk section vs RMF

  5. E5

    Significant residual risks disclosed to users, with the information needed to do so included in the IFU / label

    EN ISO 14971 §8 · Annex I §4 · Evidence: Residual risk disclosure list vs IFU

  6. E6

    MDR: clinical benefits defined and quantified as patient-relevant outcomes (e.g. survival, symptom reduction, function) with measurable parameters

    MDR Annex I §1, §8 · Art. 2(53) · Annex XIV · Evidence: CER clinical benefit definition

  7. E7

    IVDR: benefits defined as diagnostic accuracy (sensitivity, specificity, PPV, NPV) and clinical utility / public health impact

    IVDR Annex I §1, §8 · Annex XIII · MDCG 2022-2 · Evidence: PER clinical benefit definition

  8. E8

    Benefit-risk benchmarked against the state of the art and alternative treatments / diagnostic methods; alternative risk controls considered where appropriate; conclusion consistent with the intended purpose, claims and IFU

    MDR / IVDR Annex I §1 · MDCG 2020-6 · MEDDEV 2.7/1 Rev. 4 · Evidence: SOTA analysis in CER / PER

F. Risk management review and report (EN ISO 14971 §9)

0/4
  1. F1

    RM review performed before commercial release: plan implemented, overall residual risk acceptable, production / post-production methods in place

    EN ISO 14971 §9 · Evidence: RM review record

  2. F2

    RM report confirms completion of all planned activities, documents the required conclusions, addresses outstanding issues and limitations, and is reviewed and approved by the persons with assigned responsibility

    EN ISO 14971 §9 · Evidence: Signed RM report

  3. F3

    RMF revision, date and device version match the current TD revision and device configuration

    QC consistency · Evidence: RMF document control vs TD index

  4. F4

    RM report and RMF updated and re-approved when significant new information (PMS, vigilance, design or process changes) requires revision

    EN ISO 14971 §9, §10.4 · Evidence: RMF / RM report revision history

G. Production and post-production information (EN ISO 14971 §10)

0/5
  1. G1

    System for collecting and reviewing production and post-production information (complaints, NC, CAPA, vigilance, literature, similar devices, state of the art) in place

    EN ISO 14971 §10 · Annex III · Evidence: PMS procedure + PMS plan

  2. G2

    MDR: PMS / PSUR (Art. 83–86) and PMCF outputs evaluated for impact on hazards, risk estimates and acceptability

    MDR Art. 83–86 · Annex XIV Part B · Evidence: PSUR / PMS report vs RMF

  3. G3

    IVDR: PMS / PSUR (Art. 78–81) and PMPF outputs evaluated for impact on hazards, risk estimates and acceptability

    IVDR Art. 78–81 · Annex XIII Part B · Evidence: PSUR / PMS report vs RMF

  4. G4

    Trend reporting, serious-incident data and relevant field safety corrective actions (own and similar devices) reviewed for new or increased risks

    MDR Art. 87–88 / IVDR Art. 82–83 · Evidence: Vigilance and trend records

  5. G5

    Design, manufacturing, supplier and labelling changes assessed for risk impact under change control; resulting RMF changes documented and approved

    EN ISO 13485 §7.3.9 · EN ISO 14971 §10 · Evidence: Change control records

H. Traceability and consistency with the TD

0/6
  1. H1

    Each hazard traces to risk controls, verification evidence and residual risk acceptance

    EN ISO 14971 §4.5 · Evidence: Traceability matrix

  2. H2

    Warnings, precautions, contra-indications and limitations in IFU / label trace to residual risks

    MDR Annex I §23 / IVDR Annex I §20 · Evidence: IFU vs residual risk table

  3. H3

    GSPR checklist references the RMF for all applicable risk-related GSPRs

    Annex II §4 · Evidence: GSPR checklist

  4. H4

    RMF consistent with design inputs / outputs, design V&V reports, biological evaluation, sterilisation and packaging validation, and electrical safety / EMC testing, where applicable

    Annex II §3, §6 · EN ISO 10993-1 · EN ISO 11607 · IEC 60601-1 · Evidence: Cross-document consistency check (design and testing)

  5. H5

    RMF consistent with usability file, software V&V and cybersecurity documentation, PMS plan and PMS report / PSUR, and CAPA / complaint-handling records

    EN 62366-1 · EN 62304 · MDCG 2019-16 · Annex III · Evidence: Cross-document consistency check (usability, software, PMS, CAPA)

  6. H6

    Harms, side-effects and their frequencies consistent between RMF and CER / PER and SSCP / SSP

    Art. 32 (MDR) / Art. 29 (IVDR) · Evidence: RMF vs CER / PER vs SSCP / SSP

I. Reference control and QC sign-off

0/4
  1. I1

    Legal MDR / IVDR requirements, harmonised standards, non-binding guidance and internal I3CGLOBAL QC criteria kept separate

    I3CGLOBAL control · Evidence: Reference register

  2. I2

    Current harmonised standards list and latest MDCG revisions verified on the Commission website before final approval

    MDCG register · OJEU · Evidence: Dated verification record

  3. I3

    Each deficiency logged with corrective action, owner, target date and close-out sign-off

    I3CGLOBAL control · Evidence: Deficiency and corrective action log

  4. I4

    Final review signed by QA reviewer, RA specialist and clinical / performance evaluation expert

    I3CGLOBAL control · Evidence: Signed QC approval record