Sterilization TDF QC Check

IVDR Webinar

Internal · QC Department

Sterilization TD QC Check

Review the sterilization section of the MDR technical file. Mark each check OK, Issue or N/A. Issues are collected into a report that is emailed to the QC reviewer, consultant and team lead.

1. Project

2. Checklist

0/0 reviewed · 0 issues

Select the sterilization method above to show only the relevant method section.

Common NB findings: check these first

0/9
  1. N1

    Product family / worst-case rationale present and covers all variants

    Common NB finding · see B1, B2 · Evidence: Product family assessment + variant matrix

  2. N2

    Validation done on the current device and packaging revision, or change assessment provided

    Common NB finding · see B3, I4 · Evidence: Revision comparison / change assessment

  3. N3

    EO residuals tested at the correct contact category and at more than one time point

    Common NB finding · see D9, D10, L1 · Evidence: Residuals report + dissipation curve

  4. N4

    Pyrogenicity covers material-mediated pyrogens, not endotoxin only

    Common NB finding · see G5–G7 · Evidence: Rabbit test or BEP rationale

  5. N5

    Real-time aging study running alongside accelerated aging, with plan to confirm

    Common NB finding · see H7, H8 · Evidence: Real-time protocol + interim results

  6. N6

    Current standard editions used, or gap assessment for superseded editions

    Common NB finding · see A8, M5 · Evidence: Standards list / gap assessment

  7. N7

    Sterilizer's ISO 13485 certificate valid and quality agreement in place

    Common NB finding · see I5, I6 · Evidence: Certificate copy + signed agreement

  8. N8

    Latest requalification / dose audit provided

    Common NB finding · see E8, I3 · Evidence: Latest requalification or dose audit report

  9. N9

    Label shelf life does not exceed validated shelf life

    Common NB finding · see H12, J5 · Evidence: Label vs shelf-life report

A. Regulatory basis and TD structure

0/10
  1. A1

    Sterilization section exists as a clearly indexed part of the TD, not scattered across V&V

    Annex II §6.2(e) · Evidence: TD index pointing to sterilization, packaging, environment and shelf-life reports

  2. A2

    TD describes the method used, including validation reports for packaging, sterilization and maintenance of sterility

    Annex II §6.2(e) · Evidence: Method summary + validation report numbers

  3. A3

    Validation report addresses bioburden, pyrogen and (if applicable) sterilant residues

    Annex II §6.2(e) · Evidence: Each of the three explicitly covered or justified NA

  4. A4

    Description of environmental conditions for relevant manufacturing steps is included

    Annex II §6.2(e), Annex I §11.7 · Evidence: Cleanroom class, monitoring summary

  5. A5

    GSPR checklist rows 11.4, 11.5, 11.6, 11.7 and 23.3 marked applicable, with standard and document reference per row

    Annex I §11, §23 · Evidence: GSPR checklist with traceable references

  6. A6

    GSPR 11.4: packaging maintains device characteristics and sterility during transport and storage

    Annex I §11.4 · Evidence: Packaging validation + transport simulation

  7. A7

    GSPR 11.5/11.6: designed, manufactured, packaged and sterilized by validated methods; sterility maintained until opening

    Annex I §11.5, §11.6 · Evidence: Validation + shelf-life reports

  8. A8

    List of applied standards with full edition (year + amendments); harmonised status checked; deviations from harmonised editions justified

    Art. 8, Annex II §4 · Evidence: Standards list / gap assessment

  9. A9

    Document numbers, revisions and dates in TD match the attached reports

    QC consistency · Evidence: Cross-check table

  10. A10

    Reports are signed/approved and in English (or a language accepted by the NB)

    NB requirement · Evidence: Approved copies

B. Product definition, families and method rationale

0/7
  1. B1

    All models/sizes/variants in scope of the certificate are covered by the sterilization validation

    Annex II §1.1 · Evidence: Variant-to-validation matrix

  2. B2

    Product family / grouping rationale documented; worst-case (master) product identified with justification (size, mass, density, lumens, mated surfaces, bioburden)

    ISO 11135 / ISO 11137-1 product definition; AAMI TIR35 for radiation families · Evidence: Product family assessment

  3. B3

    Product adoption into an existing validated cycle/dose assessed (new or changed devices)

    ISO 11135 Annex E; ISO 11137-1 · Evidence: Adoption / equivalence report

  4. B4

    Rationale for choice of sterilization method; material compatibility demonstrated

    ISO 14937 §5; ISO 11137-1 §8.1 · Evidence: Compatibility / max-dose or max-exposure functional testing

  5. B5

    Required SAL stated (10⁻⁶ for 'STERILE' claim); any other SAL justified by risk

    EN 556-1 · Evidence: Stated SAL + demonstration in validation

  6. B6

    Device performance and safety verified after sterilization (incl. worst-case: max dose / double cycle / re-sterilization if allowed)

    Annex I §1, §11.6 · Evidence: Functional test reports post-sterilization

  7. B7

    Re-sterilization by manufacturer (rework) allowed? Number of cycles validated and reflected in procedures

    ISO 11135 / ISO 11137 · Evidence: Re-sterilization validation or prohibition in QMS

C. Manufacturing environment and contamination control

0/6
  1. C1

    Cleanroom / controlled-environment classification for assembly and primary packaging stated and justified

    ISO 14644-1; Annex I §11.7 · Evidence: Classification certificate, area layout

  2. C2

    Environmental monitoring programme: particulates, viable air/surface, alert/action limits, frequency, trending

    ISO 14644-2; EN 17141 · Evidence: EM procedure + recent trend summary

  3. C3

    Personnel gowning, hygiene and training controls in place

    ISO 13485 §6.4.2 · Evidence: Procedure reference

  4. C4

    Process water / compressed air / cleaning agents controlled where they contact product

    ISO 13485 §6.4; ISO 19227 (implant cleanliness) · Evidence: Specs + monitoring

  5. C5

    Cleaning validation for devices with machining, oils, residues (esp. implants)

    ISO 19227; ASTM F2459 · Evidence: Cleaning validation report

  6. C6

    Supplier-sterilized or supplier-cleaned components controlled

    ISO 13485 §7.4 · Evidence: Supplier agreements, incoming specs

D. Ethylene oxide (EO) validation

0/13
  1. D1

    Validation protocol approved before execution; acceptance criteria predefined

    ISO 11135 §9 · Evidence: Signed protocol

  2. D2

    Installation and operational qualification (IQ/OQ) of sterilizer and aeration equipment

    ISO 11135 §9.2–9.3 · Evidence: IQ/OQ reports, calibration records

  3. D3

    Approach stated (overkill half-cycle, BI/bioburden, or bioburden-based) and justified

    ISO 11135 Annex A/B · Evidence: Validation rationale

  4. D4

    Microbiological PQ: typically 3 half-cycles (or fractional cycles) + full cycle; BI/PCD placement at worst-case locations

    ISO 11135 §9.5 · Evidence: MPQ report with BI map

  5. D5

    PCD shown to be equal or more resistant than product (internal vs external PCD comparison)

    ISO 11135 §9.5; ISO 11138-2 · Evidence: PCD resistance study

  6. D6

    Physical PQ: temperature/humidity distribution, EO concentration, pressure profiles, load configuration and density defined

    ISO 11135 §9.4 · Evidence: PPQ data, load patterns

  7. D7

    BI certificates: B. atrophaeus, population, D-value, lot traceable

    ISO 11138-1/-2 · Evidence: BI CoA

  8. D8

    Sterility tests of product exposed to fractional/half cycles performed with method validation (bacteriostasis/fungistasis)

    ISO 11737-2 · Evidence: Test reports + suitability

  9. D9

    EO and ECH residuals tested on worst-case product, at correct exposure category (limited / prolonged / permanent contact)

    ISO 10993-7 · Evidence: Residuals report vs TCL/TL limits

  10. D10

    For implants (permanent contact): average daily dose and cumulative limits met; aeration time validated

    ISO 10993-7 §4.3 · Evidence: Dissipation curve + release aeration time

  11. D11

    Residual limits also assessed for neonates / special populations if in intended use

    ISO 10993-7 · Evidence: Justification in report

  12. D12

    Process specification (cycle parameters with tolerances) issued and matches validated cycle

    ISO 11135 §9.6 · Evidence: Process spec document

  13. D13

    Parametric release (if claimed) justified and validated

    ISO 11135 §10.2 · Evidence: Parametric release validation

E. Radiation (gamma / e-beam / X-ray) validation

0/8
  1. E1

    Maximum acceptable dose established; product and packaging functional after max dose (incl. aged samples)

    ISO 11137-1 §8.1 · Evidence: Max-dose study

  2. E2

    Sterilization dose established by recognised method: Method 1, Method 2, VDmax²⁵ / VDmax¹⁵ (ISO 11137-2) or VDmax^SD (ISO/TS 13004), or 25 kGy substantiation

    ISO 11137-2; ISO/TS 13004 · Evidence: Dose setting report

  3. E3

    Bioburden on 3 batches (10 samples/batch typical) with validated recovery efficiency used in dose setting

    ISO 11737-1; ISO 11137-2 · Evidence: Bioburden + recovery report

  4. E4

    Verification dose experiment: SIP justified, positives within acceptance

    ISO 11137-2 · Evidence: VD experiment results

  5. E5

    Dose mapping (OQ/PQ) for the defined loading pattern; Dmin/Dmax locations and ratio to routine monitoring position

    ISO 11137-1 §9; ISO 11137-3 · Evidence: Dose mapping report

  6. E6

    Dosimetry system calibrated, traceable to national standard; uncertainty considered

    ISO 11137-3; ISO/ASTM 51261 · Evidence: Calibration certificate

  7. E7

    Change of irradiator / source / e-beam ↔ gamma transfer assessed (re-mapping)

    ISO 11137-1 §12.5 · Evidence: Transfer / re-mapping report

  8. E8

    Quarterly (or justified frequency) dose audits defined; latest audit results and any augmentation

    ISO 11137-1 §12.1; ISO 11137-2 · Evidence: Recent dose audit reports

F. Moist heat, other methods and user-processed devices

0/7
  1. F1

    Moist heat: cycle (temp/time/pressure), air removal, load, heat penetration and distribution studies; BI challenge (G. stearothermophilus)

    ISO 17665 · Evidence: IQ/OQ/PQ reports

  2. F2

    VH₂O₂: lumen/material claims, load limits, residual H₂O₂ addressed

    ISO 22441; ISO 14937 · Evidence: Validation + residuals

  3. F3

    Dry heat, LTSF or novel method: validated under ISO 14937 or method-specific standard; sterilant characterisation and residuals

    ISO 20857; ISO 25424; ISO 14937 · Evidence: Validation report

  4. F4

    Aseptic processing (if terminal sterilization not possible) justified and media fills performed

    ISO 13408 series · Evidence: Justification + media fill reports

  5. F5

    Devices supplied non-sterile and sterilized by user: at least one validated method in IFU (cleaning, disinfection, drying, sterilization parameters)

    Annex I §23.4(n); EN ISO 17664-1 · Evidence: Reprocessing validation report

  6. F6

    Reusable devices: number of reprocessing cycles validated and any limit stated in IFU

    Annex I §23.4(n); ISO 17664-1 · Evidence: Lifetime / cycle study

  7. F7

    Non-sterile devices needing user sterilization: packaging minimises contamination risk

    Annex I §11.8 · Evidence: Packaging spec

G. Bioburden, sterility and pyrogen testing

0/9
  1. G1

    Bioburden method validated: recovery efficiency (repetitive or spiked), correction factor applied

    EN ISO 11737-1 · Evidence: Method validation report

  2. G2

    Routine bioburden monitoring frequency, alert/action limits and trend data

    ISO 11737-1 · Evidence: Trend summary (≥ last 12 months)

  3. G3

    Bioburden characterised (aerobes, fungi; spore-formers where relevant to method)

    ISO 11737-1 · Evidence: Characterisation data

  4. G4

    Sterility test method suitability (bacteriostasis/fungistasis) done for each product family

    EN ISO 11737-2 · Evidence: B/F test report

  5. G5

    Bacterial endotoxin test (LAL/rFC) validated: inhibition/enhancement, extraction method

    Ph. Eur. 2.6.14; USP <85>; ANSI/AAMI ST72 · Evidence: BET validation report

  6. G6

    Endotoxin limit justified: commonly 20 EU/device; 2.15 EU/device for CSF/intrathecal contact

    USP <161>; AAMI ST72 · Evidence: Limit rationale + results

  7. G7

    Material-mediated pyrogenicity addressed (rabbit test or justification)

    ISO 10993-11 · Evidence: Test report or BEP rationale

  8. G8

    Endotoxin routine testing frequency defined (batch or justified sampling plan)

    ANSI/AAMI ST72 · Evidence: Sampling plan

  9. G9

    Test labs accredited (ISO/IEC 17025) or qualified under supplier controls

    ISO 13485 §7.4 · Evidence: Accreditation scope

H. Sterile barrier system, packaging and shelf life

0/12
  1. H1

    Sterile barrier system (SBS) and protective packaging fully described: materials, configuration, suppliers

    EN ISO 11607-1 · Evidence: Packaging spec + drawings

  2. H2

    SBS material compatible with sterilization method (porosity for EO/steam; radiation stability)

    ISO 11607-1 §5 · Evidence: Material qualification

  3. H3

    Microbial barrier properties of porous materials demonstrated

    ISO 11607-1 §5.2; ASTM F1608 / F2638 · Evidence: Supplier data or test

  4. H4

    Sealing process validated (IQ/OQ/PQ): temp, pressure, dwell; seal strength and integrity at process limits

    EN ISO 11607-2 · Evidence: Seal validation report

  5. H5

    Package integrity tests: seal strength (ASTM F88), dye penetration (F1929) or bubble (F2096), visual (F1886)

    ISO 11607-1 Annex B · Evidence: Test reports

  6. H6

    Transport / distribution simulation on worst-case configuration after sterilization

    ASTM D4169 / ISTA · Evidence: Distribution test report

  7. H7

    Accelerated aging: Q10, temperature, humidity justified; AAF calculation correct

    ASTM F1980 · Evidence: Aging protocol + calculation

  8. H8

    Real-time aging study initiated in parallel; status and plan to confirm the claimed shelf life

    ASTM F1980; ISO 11607-1 §8 · Evidence: Real-time protocol + interim results

  9. H9

    Device (not only packaging) stability over shelf life verified: material degradation, functional performance

    Annex I §1, §6 · Evidence: Device stability report

  10. H10

    Aged samples were sterilized at worst-case (max dose / multiple cycles) before aging

    ISO 11607-1 · Evidence: Sample history in protocol

  11. H11

    Aseptic presentation / usability of opening evaluated

    ISO 11607-1 §7; IEC 62366-1 · Evidence: Usability evaluation

  12. H12

    Shelf life claim on label = shelf life validated (no extrapolation beyond data)

    Annex I §23.3(h) · Evidence: Label vs report match

I. Routine control, release, revalidation and contract sterilizer

0/8
  1. I1

    Routine monitoring and product release criteria defined (BI/PCD results, parameters, dosimeter readings, residuals)

    ISO 11135 §10; ISO 11137-1 §10 · Evidence: Release procedure

  2. I2

    Sample batch record / certificate of processing provided

    ISO 11135; ISO 11137-1 · Evidence: Example CoP

  3. I3

    Requalification / revalidation frequency defined (e.g., EO annual review; radiation dose audits) and latest requalification attached

    ISO 11135 §12; ISO 11137-1 §12 · Evidence: Latest requalification report

  4. I4

    Change control triggers for re-validation: product, packaging, load, equipment, site, supplier

    ISO 13485 §7.5.6 · Evidence: Change control procedure

  5. I5

    Contract sterilizer identified (name, address); ISO 13485 certificate valid and covering sterilization

    Annex IX §2.2; ISO 13485 §4.1.5 · Evidence: Certificate copy

  6. I6

    Quality / technical agreement with sterilizer defines responsibilities (validation, release, deviations, notification of changes)

    ISO 13485 §7.4 · Evidence: Signed agreement

  7. I7

    Sterilizer listed as critical supplier and subject to NB unannounced audit provisions

    Annex IX §3.4 · Evidence: Supplier list entry

  8. I8

    Non-conforming cycle handling and deviations procedure in place; history of deviations reviewed

    ISO 13485 §8.3 · Evidence: Deviation log summary

J. Labelling and IFU

0/8
  1. J1

    Label indicates sterile state and sterilization method (ISO 15223-1 symbol, e.g. STERILE EO / STERILE R)

    Annex I §23.2(n); ISO 15223-1 · Evidence: Label artwork

  2. J2

    Sterile packaging bears: sterile identification, declaration of sterile condition, method, manufacturer, device description, manufacturing date, use-by date

    Annex I §23.3(a)–(h) · Evidence: Sterile pack artwork

  3. J3

    Instruction to check IFU for what to do if sterile packaging is damaged or unintentionally opened before use

    Annex I §23.3(i) · Evidence: Label / IFU text

  4. J4

    'Do not use if package is damaged' symbol and single-use / do-not-resterilize symbols where applicable

    ISO 15223-1; Annex I §23.2(i) · Evidence: Artwork

  5. J5

    Use-by date matches validated shelf life (H12)

    Annex I §23.2(j) · Evidence: Cross-check

  6. J6

    IFU storage and handling conditions match those used in shelf-life and transport studies

    Annex I §23.4(f) · Evidence: IFU vs study conditions

  7. J7

    Single-use devices: IFU states known risks of reuse/resterilization

    Annex I §23.4(p) · Evidence: IFU text

  8. J8

    Reusable devices: validated reprocessing instructions and cycle limit in IFU (link to F5/F6)

    Annex I §23.4(n) · Evidence: IFU text

K. Links to risk management, biocompatibility, clinical and PMS

0/6
  1. K1

    Risk file covers sterility-related hazards: non-sterility, pyrogens, residuals, packaging breach, shelf-life expiry, material degradation

    EN ISO 14971; Annex I §3 · Evidence: Risk analysis rows + controls traced to validation reports

  2. K2

    Biological evaluation performed on final, sterilized (and aged where relevant) device

    EN ISO 10993-1 · Evidence: BEP/BER statement on test article

  3. K3

    Sterilant residuals and endotoxin conclusions referenced in the BER

    ISO 10993-1, -7, -11 · Evidence: BER cross-reference

  4. K4

    CER / literature on infection-related complications considered; sterility claims consistent with clinical data

    Annex XIV; MDCG 2020-6 · Evidence: CER section reference

  5. K5

    PMS plan / PSUR trend non-sterility complaints, infections, packaging failures

    Art. 83–86 · Evidence: PMS plan, complaint trend

  6. K6

    PMCF addresses long-term implant infection where relevant

    Annex XIV Part B · Evidence: PMCF plan

L. Implantable / Class III specifics, common NB findings and sign-off

0/5
  1. L1

    Permanent-contact EO residual limits applied (not limited/prolonged)

    ISO 10993-7 · Evidence: Residuals report category

  2. L2

    Endotoxin tested per batch or on justified sampling plan for implants

    ANSI/AAMI ST72 · Evidence: Release records

  3. L3

    Absorbable / polymer implants: radiation or EO effect on degradation profile and mechanical properties verified over shelf life

    Annex I §10.1 · Evidence: Post-sterilization + aged mechanical / degradation data

  4. L4

    Implant cleanliness (particulates, residual processing aids) demonstrated before sterilization

    ISO 19227 · Evidence: Cleanliness report

  5. L5

    Implant card / patient information does not contradict sterile-state information

    Art. 18 · Evidence: Implant card review

M. General document quality

0/5
  1. M1

    Overall quality of editing (language, formatting, consistency, no leftover template text)

    General QC · Evidence: Full TD read-through

  2. M2

    Manufacturer name and address are captured correctly throughout

    Annex II §1.1; Art. 10 · Evidence: TD, labels, IFU, DoC and reports cross-check

  3. M3

    Procedures / protocols are correctly identified and numbered

    General QC · Evidence: Document list vs references in TD

  4. M4

    Records are correctly captured, identified and numbered

    General QC · Evidence: Record IDs vs references in TD

  5. M5

    Latest and prevailing harmonised and non-harmonised standards used throughout

    Art. 8; Annex II §4 · Evidence: Standards list vs OJEU / current editions

3. QC report

Not started

Progress is saved in this browser only.