Class Ir Reusable Surgical Instruments under EU MDR

Class Ir, also written Class 1r, covers reusable surgical instruments such as scalpel handles, forceps, scissors and retractors. Rule 6 of EU MDR Annex VIII places them in Class I. Article 52(7) adds a Notified Body review of reuse: cleaning, disinfection, sterilisation, maintenance, functional testing and the instructions for use. The category did not exist under the MDD, where these instruments were self-declared Class I.

Classification rules

Which Rule Makes an Instrument Class Ir?

Rule 6 covers surgically invasive devices for transient use (normally less than 60 minutes of continuous use). These are normally Class IIa, but Rule 6 makes an exception: reusable surgical instruments are Class I.

Article 2(8) defines a reusable surgical instrument as one that is:

  • intended for surgical use in cutting, drilling, sawing, scratching, scraping, clamping, retracting, clipping or similar procedures,
  • not connected to an active device, and
  • intended by the manufacturer to be reused after appropriate procedures such as cleaning, disinfection and sterilisation.

Stricter rules still win: if a reusable instrument is intended to contact the heart, the central circulatory system or the central nervous system directly, Rule 6 makes it Class III, and the Class Ir route does not apply.

Examples

What Are Examples of Class Ir Instruments?

These are typical reusable surgical instruments. The intended purpose and the absence of any active connection decide the class.

Instrument (typical)Surgical action
Scalpel handlesCutting (with a single-use blade classified separately)
Surgical scissorsCutting
Forceps and needle holdersClamping and holding
RetractorsRetracting tissue
Curettes and raspsScraping
Manual bone drills and sawsDrilling and sawing, without power
Haemostatic clampsClamping

CE marking route

How Do You CE Mark Class Ir Instruments?

Article 52(7)(c) lets the manufacturer use Annex IX Chapters I and III, or Annex XI Part A. The Notified Body assesses only the aspects of reuse. Eight steps take a reusable instrument from classification to the CE mark.

  1. Confirm Class IrCheck the Article 2(8) definition and rule out Class III contact with the heart, circulation or nervous system.
  2. Write the reprocessing instructionsDraft cleaning, disinfection, drying, inspection, maintenance, packaging and sterilisation steps in line with EN ISO 17664-1.
  3. Validate reprocessingValidate manual and automated cleaning, disinfection and sterilisation with worst-case soils and instruments.
  4. Set the reuse limitTest function and materials over repeated cycles and state the number of reuses or the end-of-life criteria.
  5. Complete the technical documentationPrepare the Annex II and III files, the clinical evaluation, risk management and the GSPR checklist.
  6. Apply to a Notified BodyChoose a Notified Body designated for reusable surgical instruments and the conformity assessment annex.
  7. Pass the reuse assessmentThe Notified Body reviews validation and the instructions for use, and audits the related QMS processes.
  8. Sign the declaration and affix the CE markAdd the Notified Body number next to the CE mark. Apply UDI; Class I reusable devices must carry the UDI on the device itself from 26 May 2027.

Reprocessing evidence

What Reprocessing Validation Does Class Ir Need?

Validation must show that the reprocessing steps in your instructions for use work every time, and that reuse does not reduce safety or performance.

AreaTypical referenceWhat to show
Reprocessing informationEN ISO 17664-1Complete, validated instructions that hospitals can follow.
CleaningEN ISO 15883 seriesManual and automated cleaning remove worst-case soil to set limits.
DisinfectionEN ISO 15883 seriesThe stated disinfection method reaches the required level.
SterilisationEN ISO 17665The recommended steam cycle sterilises the instrument in its packaging. Other methods follow their own standard, for example EN ISO 11135 for ethylene oxide.
DurabilityManufacturer’s protocolFunction, corrosion resistance and material integrity after the claimed number of cycles.

Robustness

Test the worst case: the hardest-to-clean design features and the heaviest soil.

Repeatability

Run several cycles to show the result is consistent.

Reproducibility

Show that results hold across operators, equipment and test runs.

After CE marking. Your obligations continue after the certificate is issued.

ObligationMDR referenceNote
Post-market surveillance plan and PMS reportArticles 83 to 85Update the PMS report when necessary.
Vigilance: serious incidents and field safety corrective actionsArticle 87Report through the vigilance system.
Notified Body surveillance of reprocessing processesAnnex IX or Annex XITell the Notified Body about significant changes before you implement them.
EU authorised representative (manufacturers outside the EU)Article 11Named on the label and registered in EUDAMED.
Person responsible for regulatory complianceArticle 15Required for every manufacturer; micro and small enterprises may use an external PRRC.
Technical documentation retentionArticle 10(8)At least 10 years after the last device is placed on the market; 15 years for implantable devices.

UDI

When Must Class Ir Instruments Carry a UDI?

Reusable surgical instruments must carry the UDI carrier on the device itself, because the label and packaging are discarded before each reuse (MDR Annex VI Part C, section 4.10). For Class I devices the UDI on the label has applied since 26 May 2025, and direct marking of reusable Class I devices applies from 26 May 2027 under Article 123(3)(g). Register the Basic UDI-DI and UDI-DI in EUDAMED once the relevant module is mandatory.

Device classUDI on the labelDirect marking of reusable devices
Class I, including Class Ir26 May 202526 May 2027
Class IIa and IIb26 May 202326 May 2025
Class III and implantable26 May 202126 May 2023

Source: European Commission UDI helpdesk, UDI carrier deadlines. See our UDI for medical devices guide for the assignment and labelling steps.

Borderline cases

What Are Common Class Ir Mistakes?

  • Calling powered tools Class Ir. Instruments connected to an active device are not reusable surgical instruments under Article 2(8).
  • Single-use parts. A single-use scalpel blade is surgically invasive for transient use and is usually Class IIa, separate from the reusable handle.
  • Heart or CNS contact. Reusable instruments for direct contact with the heart, central circulation or central nervous system are Class III.
  • No reuse limit. Notified Bodies expect a validated number of cycles or clear end-of-life criteria.
  • IFU not matching the validation. Every cleaning and sterilisation parameter in the IFU must be the one you validated.
  • Treating the certificate as a full QMS certificate. The Class Ir certificate covers the aspects of reuse only; it does not certify the rest of the quality management system.
  • Relying on the legacy transition. The extended transition under Regulation (EU) 2023/607 runs until 31 December 2028 only for manufacturers who met its 2024 conditions.

Compare

How Does Class Ir Compare with Other Classes?

ClassNotified Body scope
Class INone, self-declared
Class IsSterile conditions only
Class ImMetrological requirements only
Class IrReuse aspects only
Class IIaFull QMS and technical documentation sampling, for example single-use surgical instruments

An instrument that is supplied sterile and intended for reuse needs both aspects assessed: the sterile condition under Article 52(7)(a) and the reuse aspects under Article 52(7)(c), usually by the same Notified Body in one application.

See every class side by side on Class I vs IIa vs IIb vs III.

Work with I3CGLOBAL

Certify Your Reusable Surgical Instruments

I3CGLOBAL has completed 300+ CE certifications, with offices in India, Europe and the USA. We help instrument manufacturers prepare Class Ir files and reprocessing evidence.

Request your proposal

Share your instrument range and we will prepare a Class Ir proposal.

Request an MDR QuoteContact Us

FAQs

Class Ir Reusable Surgical Instruments: Frequently Asked Questions

Is Class 1r the same as Class Ir?

Yes. Class Ir and Class 1r are the same category: reusable surgical instruments classified as Class I under Rule 6 of EU MDR Annex VIII. The small r stands for reusable.

Does a Class Ir instrument need a Notified Body?

Yes. Under Article 52(7)(c), a Notified Body must assess the aspects of reuse, in particular cleaning, disinfection, sterilisation, maintenance and functional testing, and the related instructions for use. This was not required under the old MDD.

What is a reusable surgical instrument under the MDR?

Article 2(8) defines it as an instrument for surgical use in cutting, drilling, sawing, scratching, scraping, clamping, retracting, clipping or similar procedures, without a connection to an active device, and intended to be reused after appropriate reprocessing.

Can legacy Class Ir instruments still be sold under an MDD declaration?

Only if the manufacturer met the conditions of Regulation (EU) 2023/607, including a Notified Body application by 26 May 2024 and a written agreement by 26 September 2024. Those devices can be placed on the market until 31 December 2028. No new legacy applications are possible, so a new Class Ir instrument must be certified under the MDR.

Is a powered surgical drill Class Ir?

No. An instrument connected to an active device does not meet the definition of a reusable surgical instrument. Powered surgical tools are classified under the rules for active or surgically invasive devices, usually in a higher class.

What causes most Class Ir Notified Body findings?

Most findings concern weak reprocessing validation: missing worst-case soils, too few cycles, no limit on the number of reuses, or instructions for use that do not match the validated cleaning and sterilisation steps.

When must Class Ir instruments carry the UDI directly on the device?

From 26 May 2027. Reusable devices that are cleaned, disinfected or sterilised between uses must carry the UDI carrier on the device itself. For Class I reusable devices the deadline under Article 123(3)(g) is 26 May 2027; the UDI on the label has applied since 26 May 2025.

What certificate does the Notified Body issue for Class Ir?

An EU quality management system certificate under Annex IX, or an EU quality assurance certificate under Annex XI Part A. Either certificate is limited to the aspects of reuse named in Article 52(7)(c). It does not certify the rest of the quality management system, so many manufacturers also hold a separate ISO 13485 certificate.

Is a clinical evaluation needed for a Class Ir instrument?

Yes. Article 61 requires a clinical evaluation for every device class. For well-established reusable instruments it is usually based on published literature and equivalence. Article 61(10) allows a justification based on non-clinical data where clinical data are not considered appropriate, and that justification must be documented in the technical file.

What if an instrument is reusable and also supplied sterile?

Both aspects are assessed. Article 52(7) lists sterile condition and reuse as separate Notified Body scopes. An instrument placed on the market sterile and intended for reuse needs the sterile aspects assessed under Article 52(7)(a) and the reuse aspects under Article 52(7)(c), usually by the same Notified Body in one application.

How long does Class Ir certification take?

It depends on two things: how long the reprocessing validation testing takes, and the Notified Body’s scheduling. Validation of cleaning, disinfection, sterilisation and durability across the claimed number of cycles is usually the longest step, so start it before the Notified Body application. I3CGLOBAL gives a project-specific timeline in the proposal.